Corrosion-resistant and high-temperature resistant paddle agitator with top-entry design
Contact Info
- No.4285 Shendu Highway, Pujiang Town, 200000
- 胡工
- 18001633556
- 15000845544@163.com
Top-entry agitatorCorrosion-resistant and high-temperature resistant paddle agitator
1. Product Introduction
Top-entry agitatorCorrosion-resistant and high-temperature resistant paddle agitator can be applied in fields such as biopharmaceuticals, biochemistry, food, nanomaterials, coatings, adhesives, daily chemicals, printing and dyeing, petrochemicals, papermaking chemistry, etc. This digital display electric agitator is suitable for mixing processes such as dispersion and agitation of multi-viscosity systems in laboratories.
2.Add new functions for your laboratory
1. Digital display
2. Powerful, vertically elongated and humanized design
The motor performance and technology have been improved based on the reliable YK120 series, with stable motor power
3. Two speed gears (high and low speed), ranging from 60 to 2000rpm.
4. Penetration stirring paddle (only when fixed).
Equipment Parameters | Numerical Value |
Power Supply | 220-240/100-115 V |
Motor Power | 120W |
Agitation Volume (H2O) | 20 L |
Motor Input Power | 70 W |
Motor Output Power | 35 W |
Speed Display | LED liquid crystal display |
Speed Range | 60 - 2000 rpm |
Processing Viscosity | 10000CP |
Stirring Shaft Output Power | 26 W |
Allowable Continuous Running Time | 100 % |
Stirring Shaft Torque | 150 Ncm |
Stirring Shaft Torque at 60 rpm (overload) | 300 Ncm |
Stirring Shaft Torque at 100 rpm | 150 Ncm |
Stirring Shaft Torque at 1000 rpm | 24 Ncm |
Speed Range I (50 Hz) | 60 - 500 rpm |
Speed Range II (50 Hz) | 240 - 2000 rpm |
Speed Range I (60 Hz) | 72 - 600 rpm |
Speed Range II (60 Hz) | 288 - 2400 rpm |
Speed Control | Stepless |
Stirring Paddle Fixation | Rotary chuck |
Clamping Range Diameter of Rotary Chuck | 0.5-10 mm |
Hollow Shaft, Inner Diameter | 10.5 mm |
Hollow Shaft (Insertable in Stopped State) | Yes |
Support Fixation | Extended arm chuck fixation |
Extended Arm Diameter | 13 mm |
Extended Arm Length | 160 mm |
Rated Torque | 1.5 Nm |
Allowable Ambient Temperature | 5 - 40 °C |
Allowable Relative Humidity | 80 % |
Protection Level per DIN EN 60529 | IP 20 |
Overall Dimensions | 880 x 294 x 212 mm |
Weight | 3.1 kg |
3. Optional Configurations
1. Four-blade propeller type --- Standard stirring head, which makes the mixed medium axially flow from top to bottom with low shear force.
2. Three-blade propeller type --- Makes the mixed medium axially flow from top to bottom, generating local shear force. Suitable for medium and high speed use.
3. Dissolving type --- Makes the mixed medium generate radial flow both upward and downward, and the formed strong turbulence and high shear force can crush particles in liquid. Suitable for medium and high speed use.
4. Centrifugal type --- Double blades expand as the speed increases, suitable for use in narrow-mouth containers. The stirring effect is similar to that of the four-blade propeller type. Suitable for medium and high speed use.
5. Fan blade type --- Generates tangential flow with small turbulence and good heat exchange performance, applied in occasions where the medium needs gentle stirring. Suitable for low and medium speed use.
6. Anchor type --- Generates tangential flow with high shear rate at the edge, preventing the mixed medium from depositing on the container wall. Suitable for low speed use. It is an ideal stirring element for medium and high viscosity fluids, applied in polymer reactions and dispersion of minerals in liquids.
7. Other special blades --- Various other styles of blades can be selected to meet your special working conditions.
4.Maintenance of Mechanical Agitator:
· 1. The bearing bears all the loads of the machine, so lubrication has a great relationship with the bearing life. It directly affects the service life and operation rate of the machine, so it is required that the injected lubricating oil be clean and well sealed. The main oil injection points of this machine are (1) rotating bearings (2) roller bearings (3) all gears (4) movable bearings and sliding surfaces.
· 2. The newly installed wheel band is prone to loosening and needs regular inspection.
· 3. Pay attention to whether the working conditions of all parts of the machine are normal.
· 4. Pay attention to checking the wear degree of easily worn parts, and replace the worn parts at any time.
· 5. The bottom frame plane of the movable device should be free of dust and other substances to avoid serious accidents when the machine encounters unbreakable materials and the movable bearing cannot move on the bottom frame. 6. If the bearing oil temperature rises, stop the machine immediately to check and eliminate the cause.
| Chemicals/Mixing Equipment/Mixers | |