Endress+Hauser (E+H) PMC131-A22F1A2K Ceramic Pressure Transmitter (0-2bar)

Product Description

Ceramic measuring element pressure sensor

Can measure gauge pressure and absolute pressure

High reliability and overload resistance

Application

Suitable for measuring absolute pressure and gauge pressure of gases, steam and liquids.

Features and Advantages

· Ceramic measuring element ensures measurement safety

· Reliability

· High stability

· Easy installation

Measuring System

· Cerabar T pressure sensor with 4...20mA output

· Power supply voltage: 11-30V, when using power supply units provided by E+H such as RN221, NX9120 or NX9121

Technical Data

General Information

Manufacturer: E+H

Instrument Name: Pressure Transmitter

Model: PMC131-A22F1A2K

Application

Measure gauge pressure and absolute pressure of gases, steam and liquids

Operation and System Design

Measuring Principle: The measured pressure generates a small deformation on the ceramic diaphragm of the sensor. The electrodes on the ceramic sensor measure the resulting change in capacitance proportional to the pressure.

Modularity: The pressure sensing element can be connected to the transmitter power supply via an optional DIN43650 A/ISO4400 plug or cable gland.

Input

Measured Variable: Gauge pressure or absolute pressure

Measuring Range: Max. 40bar (see selection table)

Output

Output Signal: 4⋯20mA

Load: R(Ω) ≤ (U-11V)/0.02A

Accuracy

Reference Conditions: Complies with DIN IEC 770

Rise Time: 20ms (T time)

Measurement Deviation (including hysteresis and repeatability): 0.5% FS

Nominal Range Ambient Temperature Effect: Zero signal change between -20⋯+85℃ (-4...+185°F), typical: 1.5%; Range change between -20⋯+85℃ (-4...+185°F), typical: 0.8%; (0.4bar⋯40bar), typical: 1.0% (0.1bar)

Long-Term Stability: 0.15%/year

Basismodell

Pressure Transmitter PMC131-A22F1A2K

Electrical Connection:

A2 ISO4400 NPT1/2 plug, IP65 NEMA4X

Process Connection:

2 ANSI MNPT1/2 FNPT1/4 thread, 304

Sensor Seal:

F FKM Viton

Additional Options:

1 Basic type

Measuring Range; Sensor Overload Limit:

A1K 0...2bar/200kPa abs; 18bar/1.8MPa