University-level equipment for detecting total energy in feed
Other Products
University equipment for detecting total energy in feed University equipment for detecting total energy in feed Hebi Biaochang Instrument Manufacturing Co., Ltd.
Process:
Turn on the computer, click the calorimeter icon on the calorimeter, switch on the power of the calorimeter, then click the water injection to even out the water temperature.
Calibration (heat capacity). Enter the same number, perform three benzoic acid tests consecutively, take the two closest values, average them, and save.
Reverse calibration (calorific value). Perform one calorific value test with benzoic acid to check if the bomb calorimeter value is around 6330 kcal. If it is close, the machine is normal and you can proceed to the next step; if the difference is large, recalibrate the heat capacity until it is close to 6330 kcal.
Basic Parameters
Temperature measurement range: 5℃~40℃
Temperature resolution: 0.0001℃
Error: ≤0.1%
Test cycle: ≤13min
Measurement range: 5MJ/kg~40MJ/kg
Accuracy: Better than GB/T384
Power supply: AC220V±10V/50Hz
Power: ≤0.5kW
Main unit dimensions (mm): 830×460×420
Main unit weight (kg): 82