ND6031 Fully Automatic Open Cup Flash Point Tester
Contact Info
- Add:武汉市东湖开发区高新四路40号, Zip: 430074
- Contact: 蔡先生
- Tel:+8613871491300
- Email:whnort@163.com
Other Products
Product Introduction
ND6031 Fully Automatic Open Cup Flash Point Tester is designed and manufactured in accordance with the requirements specified in the standard GB/T 3536 "Determination of Flash and Fire Points of Petroleum Products (Cleveland Open Cup Method)". It is suitable for determining the flash point of petroleum products, excluding fuel oil and those with an open flash point below 79°C, using the Cleveland open cup method as stipulated by this standard.
This instrument adopts microcontroller control technology, a color LCD display, a fully Chinese human-machine dialogue interface, and unmarked keyboards. It features menu prompts and guided input functions for parameters such as preset flash point temperature, sample number, atmospheric pressure, and test date. It utilizes technologies such as open, fuzzy control integrated software, and modular structural design.
With advanced design, a user-friendly interface, convenient operation, and accurate test results, this instrument is an ideal substitute for similar imported products. It can be widely used in industries such as railways, aviation, power, and petroleum, as well as in universities, research institutes, and metrology inspection departments for the detection and testing of flash and fire points of petroleum products.
Product Features
☆、This instrument features a large color LCD screen display, a fully Chinese human-machine dialogue interface, and unmarked keyboards. It offers prompt menus and guided input functions for parameters such as preset temperature, sample number, atmospheric pressure, and test date.
☆、It simulates and tracks the functional curve of temperature rise versus test time, includes Chinese error operation software prompts for modification, and provides parameter prompts such as test date and test time.
☆、Automatically corrects the influence of atmospheric pressure on the test and calculates the correction value.
☆、Differential detection with automatic system deviation correction.
☆、Automatically completes sweeping, detection, and data printing, with the test arm automatically rising and falling.
☆、Electronic ignition and forced air cooling (Note: The manual ignition button is located on the left side of the instrument case).
Technical Parameters
☆、Power supply: AC220V±10%,50Hz
☆、Temperature measurement: Range: Room temperature ~400℃;
Repeatability: ≤4℃;
Reproducibility: ≤8℃;
Resolution: 0.1℃;
Accuracy: 0.5%;
☆、Heating rate: Complies with GB/T3536 standard;
☆、Ignition method: Electronic ignition, gas flame 3mm~4mm;
☆、Ambient temperature: (10~40)℃;
☆、Relative humidity: ≤80%;
☆、Power consumption: Not more than 500W.
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 武汉诺顿whnort |
| Spec: | ND6031 |
| Stock: | 100 |
| Manufacturer: | |
| Origin: | China / Hubei / Wuhanshi |