KX-3A Spastic Muscle Low-Frequency Therapy Apparatus
Contact Info
- Add:北京市海淀区紫雀路33号院3号楼四层, Zip:
- Contact: 陈贞
- Tel:13258102986
- Email:465364378@qq.com
Other Products
KX-3A Spastic Muscle Low-Frequency Therapy DeviceProduct Overview:
1. Belongs to low-frequency pulse therapy devices;
2. Features dual-channel output (A and B) with continuously adjustable pulse cycle, pulse width, output intensity, and delay time of channel B relative to channel A;
3. Timer offers six selectable settings up to 30 minutes;
4. Includes misadjustment indicator;
5. This device can be used to treat various spastic paralysis conditions, as well as for functional electrical stimulation, electrogymnastics, and high-efficiency electroacupuncture therapy.
KX-3A Spastic Muscle Low-Frequency Therapy DeviceUsage Methods:
1. Alternating stimulation of spastic muscles and antagonistic muscles (for limb spastic paralysis);
2. Alternating stimulation of trunk muscle spasms;
3. Alternating stimulation to enhance bidirectional joint mobility;
4. Electroacupuncture therapy;
5. Unipolar motor point stimulation electrotherapy and acupoint electrotherapy (for distal limbs or small muscles).
KX-3A Spastic Muscle Low-Frequency Therapy Device Main Performance:
(1) Output waveform: Both channel A and B outputs are non-polarized bidirectional asymmetric pulses.
(2) Pulse cycle: Adjustable from 1s to 2s continuously, tolerance ±15%.
(3) Pulse width: Adjustable from 0.1ms to 0.5ms continuously, tolerance ±30%.
(4) Single pulse charge at maximum output: Should be greater than 7μC.
(5) Delay time: Channel B output pulse appears delayed relative to channel A, with delay time T1 adjustable from 0.1s to 1.5s continuously, tolerance ±15%.
(6) Output intensity: Peak pulse current Ip for both channels A and B is adjustable from 0mA to 99mA continuously. Maximum output tolerance ±10%.
(7) Timer settings: Six settings: 5, 10, 12, 20, 25, 30 minutes, tolerance ±5%.
(8) Misadjustment indicator: Activates when pulse cycle is less than or equal to delay time due to improper adjustment.
(9) Maximum output voltage: Peak output voltage should not exceed 500V when output is open-circuited.
(10) Resistance to output short-circuit and open-circuit: The device should withstand open and short circuits without performance degradation.
(11) Output upon power restoration: When output exceeds 10mA (RMS) or 10V (RMS) on a 1kΩ load, the device must not output after power interruption and restoration unless output amplitude is preset to minimum.
(12) Output adjustment and minimum output: Output amplitude should adjust continuously and uniformly, or in discrete increments no greater than 1mA or 1V. Minimum output should not exceed 2% of maximum output.
(13) Output DC component: DC component of output should be zero.
| Industry Category | Furniture |
|---|---|
| Product Category | |
| Brand: | 康达 |
| Spec: | 13258102986(微信同号) |
| Stock: | 1000 |
| Manufacturer: | |
| Origin: | China / Beijing / Haidianqu |