XKYKL-3 Cement Limestone Calcium Iron Element Analyzer
Contact Info
- Add:湖南省长沙县安沙镇安沙社区护丰组, Zip:
- Contact: 王经理
- Tel:15573135383
- Email:csxkyyq@163.com
Other Products
Main Applications
X-ray fluorescence calcium and iron element analyzers are widely used in industries such as metallurgy, mining, and building materials. In cement plants, they are primarily used to measure the percentage content of CaO and Fe2O3 in raw materials, cement, limestone, etc.; they can be connected to a ratio-based batching system to achieve automatic control, facilitating timely adjustment of raw material proportions; by detecting the Fe2O3% and CaO% in cement, the amount of mixed materials added can be estimated, helping cement enterprises produce qualified cement of a certain grade without using excessive clinker or insufficient mixed materials, thereby creating significant economic benefits for cement enterprises annually.
Main Features
1. Wide measurement range, with CaO reaching a width of 15%;
2. Fast analysis time, capable of measuring the percentage content of CaO and Fe2O3 in just 1 minute;
3. Non-destructive testing, samples can be reused;
4. No chemical reagents required, no waste emissions, no radioactive sources, low power consumption, meeting environmental and energy-saving requirements;
5. Large data storage capacity, content results, and instrument self-test data can be queried;
6. Integrated microcomputer design, compact structure, and aesthetically pleasing appearance;
7. 8-inch capacitive touch screen display, full Chinese menu prompts for operation, extremely convenient to use;
8. Precision stepper motor for automatic sample feeding, precise positioning, and easy to use;
9. RS232 communication interface, can connect to a computer or thermal printer.
10. The instrument complies with GB/T19140 "General Rules for X-ray Fluorescence Analysis of Cement".
Technical Specifications
1. Analysis range: 0.01%~100%
2. Analysis precision: SCaO≤0.09%, SFe2O3≤0.05%
3. Analysis width: max—min of each component ≤15%, working range selected through calibration of the working curve.
4. Analysis time: 30 seconds, 60 seconds, 120 seconds.
5. Operating conditions: power supply AC200V~240V; ambient temperature 0~40℃; relative humidity <85%
6. Total power consumption: <50W
7. Overall dimensions: 468mm*368mm*136mm
8. Total weight: 13.8kg
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 欣科源 |
| Spec: | XKYKL-3 |
| Stock: | |
| Origin: | China / Hunan / Changshashi |