DS-PO4200 Orthophosphate Water Quality Online Automatic Monitor
Contact Info
- Add:西安市高新区锦业二路逸翠尚府北区8号楼1单元1701, Zip: 710000
- Contact: 魏经理
- Tel:15686405979
- Email:1528126689@qq.com
Other Products
Measurement PrinciplePrinciple
Ammonium molybdate spectrophotometry. In an acidic medium, orthophosphate reacts with a chromogenic agent and a reducing agent to form a colored complex. The absorbance is measured at a specific wavelength and is proportional to the concentration of orthophosphate.
ØThe ten-way valve uses a sapphire valve core with high wear resistance, resistance to strong acids and alkalis, organic solvents, long service life, and high stability;
ØA specially designed reference method is adopted, where minor displacement of the digestion cell and minor contamination do not affect the measurement value;
ØEquipped with waste liquid separation function, enabling separate discharge of high-concentration reaction waste and low-concentration cleaning waste;
ØFeatures dynamic dilution function, automatically adjusting the dilution ratio based on the on-site set range;
ØIncludes built-in quality control functions such as spike recovery, blank testing, standard sample verification, zero-point verification, span verification, and parallel sample testing, and supports automatic, manual, and remote initiation of quality control;
ØCapable of automatic fault diagnosis and uploading/recording of abnormal information, such as component failure, over-range alarm, exceedance alarm, reagent shortage alarm, etc.;
ØEquipped with overheating protection: automatically cuts off power to the heater when the temperature is too high;
ØAfter abnormal reset or power restoration, the instrument automatically discharges residual reactants and returns to the initial working state. Features power-on alarm function, generating an alarm message when power is restored after an outage;
ØCapable of recording process logs, measurement data, and alarm information. With data recorded hourly, it can store over 5 years of data and supports quick export via USB drive;
ØFeatures three-level management permissions.
Typical ApplicationsApplication
ØOnline monitoring of orthophosphate in drinking water and surface water;
ØOnline monitoring of orthophosphate at municipal sewage outlets, pollution source discharge points, industrial wastewater outlets, secondary sedimentation tanks, and high-efficiency sedimentation tanks.
Technical ParametersSpecification
Range | 0-10 mg/L, as P (customizable) 0-30 mg/L, as PO4 (customizable) Switching between as P and as PO4 is available for operators and higher permissions |
Indication Error | ≤±5% or ≤±0.005 mg/L, whichever is greater |
Detection Limit | ≤0.01 mg/L |
Quantitation Limit | ≤0.02 mg/L |
Repeatability | ≤2% |
Resolution | 0.0001 mg/L (decimal places configurable from 2 to 4 digits) |
Measurement Cycle | 23 minutes |
Display | 7-inch touch screen |
Measurement Modes | Continuous mode, periodic mode (adjustable from 1h to 30d), fixed-point mode, controlled mode, and manual mode available |
Calibration Modes | Manual calibration and automatic calibration modes |
Maintenance Cycle | Generally once per month |
Power Supply | 220VAC±10%, 50HZ±10% |
Output | 4-20mA output |
Communication | MODBUS RS-232, RS-485 |
Environmental Conditions | Indoor environment with adjustable temperature, recommended range: +5 to 40°C; humidity ≤85% (non-condensing) |
Dimensions | 302*271*462mm (L*W*H) |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 西安东升 |
| Spec: | DS-PO4200 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |