DS-TFE200 Total Iron Water Quality Online Automatic Monitor
Contact Info
- Add:西安市高新区锦业二路逸翠尚府北区8号楼1单元1701, Zip: 710000
- Contact: 魏经理
- Tel:15686405979
- Email:1528126689@qq.com
Other Products
Measurement PrinciplePrinciple
Main FeaturesFeatures
uUses a sapphire valve core ten-way valve with high wear resistance coefficient, resistant to strong acids and alkalis, organic solvents, long service life, and high stability;
uEmploys a specially designed reference method, where minor displacement or contamination of the digestion cell does not affect the measurement value;
uFeatures waste liquid separation function, enabling separate discharge of high-concentration reaction waste and low-concentration cleaning waste;
uIncludes dynamic dilution function, automatically adjusting the dilution ratio based on the on-site set range;
uEquipped with quality control functions such as built-in spiked recovery, blank testing, standard sample verification, zero point verification, span verification, and parallel sample testing, which can be initiated automatically, manually, or remotely;
uCapable of automatic fault diagnosis and uploading/recording of abnormal information, such as component failure, over-range alarm, exceedance alarm, reagent shortage alarm, etc.;
uFeatures overheating protection: automatically cuts off power to the heater if the temperature is too high;
uAfter abnormal reset or power restoration, the instrument automatically discharges residual reactants and returns to the initial working state. Includes power-on alarm function, generating an alarm message upon restart after power loss;
uCapable of recording process logs, measurement data, and storing alarm information. With data recorded hourly, it can store over 3 years of data and allows quick export via USB drive;
uFeatures three-level management permissions;
Typical ApplicationsApplication
lOnline monitoring of iron concentration in tap water, drinking water, and surface water;
lOnline monitoring of iron concentration at municipal sewage inlets/outlets, pollution source discharge points, and industrial wastewater outlets.
Technical SpecificationsSpecification
Range | 0-20mg/L (customizable) |
Indication Error | ≤±5% |
Repeatability | ≤2% |
Resolution | 0.0001 mg/L (decimal places configurable from 2 to 4 digits) |
Display | 7-inch touch screen |
Measurement Modes | Continuous mode, periodic mode (adjustable from 1h to 30d), fixed-point mode, controlled mode, manual mode optional |
Calibration Modes | Manual calibration and automatic calibration modes |
Maintenance Cycle | Generally once per month |
Power Supply | 220VAC±10%, 50HZ±10% |
Output | 4-20mA output |
Communication Methods | MODBUS RS-232, RS-485 |
Environmental Conditions | Indoor with adjustable temperature, recommended temperature +5~40℃; humidity ≤85% (non-condensing) |
Dimensions | 302*271*462mm (L*W*H) |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 西安东升 |
| Spec: | DS-TFE200 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |