DS-WQ2000 Pipeline Water Quality Monitoring System
Contact Info
- Add:西安市高新区锦业二路逸翠尚府北区8号楼1单元1701, Zip: 710000
- Contact: 魏经理
- Tel:15686405979
- Email:1528126689@qq.com
Other Products
Product Introduction
The pipe network water quality monitoring system is an integrated system that combines water quality monitoring sensors, data processing units, and internal water flow pipeline units. It can continuously measure temperature, pH, turbidity, disinfectant indicators, and can be expanded to measure water quality parameters such as conductivity and ORP.
The disinfectant indicators use the DPD colorimetric method, with optional residual chlorine and total chlorine disinfectant indicators; its cuvette has an automatic cleaning function, greatly reducing maintenance and improving the stability of measurement data.
Additionally, the disinfectant indicators can use the constant voltage method, with optional residual chlorine and chlorine dioxide disinfectant indicators; its electrode structure is simple, easy to clean and replace, requiring no membrane or reagent replacement, making maintenance straightforward.
Turbidity is measured using the EPA principle of 90° scattered light technology, employing an LED light source with low power consumption and a long lifespan.
Features
lIntegrated design, small footprint;
lHigh integration of internal wiring and water circuits, users only need to connect power and water inlets/outlets, enabling easy on-site installation;
lInternal water circuit design ensures stable water flow during measurement, guaranteeing accurate monitoring data;
lDPD colorimetric analyzer with automatic cleaning function, reducing maintenance;
lEquipped with a 7-inch color touchscreen, simple operation interface, user-friendly, reliable system, and low maintenance;
lData storage and curve display functions to track changes in water quality dynamics;
lAutomatic fault diagnosis function, with abnormal information recording and uploading, such as component failure, over-range alarms,超标报警, etc.;
lThree-level management权限;
lHistorical data download and curve display for直观准确地了解 and tracking changes in water parameters over a certain period;
lOptional wireless data transmission module.
Typical Applications
lWater quality monitoring in自来水 water plants and供水 pipe networks;
lSecondary water supply quality testing in residential areas;
lReal-time water quality monitoring in swimming pools, SPAs, etc.;
lWater quality monitoring in rural自来水 systems.
Main Unit PhysicalSpecifications
Display Output | 7-inch touchscreen with strong LED backlight for operation under direct sunlight |
Power Supply | 220VAC, 50Hz/60 Hz |
Output | 8 channels of 4-20mA output, programmatically set response parameters and values |
Relay | Four relays, programmatically set response parameters and values |
Communication Protocol | Standard MODBUS RS485 communication function for real-time transmission of measured values Optional wireless communication methods: 4G: UDP/TCP NB: UDP |
Storage Temperature | 0~60℃ |
Operating Temperature | 0~45℃ (non-freezing) |
Protection Level | IP54 |
Dimensions | 340mm×450mm×900mm (L×W×H) |
Monitoring Parameter Specifications
Residual Chlorine, Total Chlorine | Principle | DPDStandard Colorimetric Method |
Range | 0-10ppm | |
Resolution | 0.01ppm | |
Accuracy | 0~5ppm: ±5% of reading or ±0.03ppm Cl2, whichever is greater;5~10ppm: ±10% of reading | |
Residual Chlorine,Chlorine Dioxide | Principle | Constant Voltage Method |
Range | 0-20ppm | |
Resolution | 0.01ppm | |
Sensitivity | 78%~155% | |
pH | Principle | Glass Electrode Method |
Range | 0-14 pH | |
Resolution | 0.01 pH | |
Accuracy | ±0.1 pH | |
ORP | Principle | Glass Electrode Method |
Range | -2000mV~2000mV | |
Resolution | 1mV | |
Accuracy | ±5mV | |
Conductivity | Principle | Conductivity Cell Method |
Range | 10μS/cm-5000 μS/cm(K=1) | |
Resolution | 0.1μS/cm~0.01mS/cm(depending on range) | |
Accuracy | ±1% or 1μS/cm, whichever is greater | |
Turbidity | Principle | Light Scattering |
Range | 0-100NTU | |
Resolution | 0.001~0.1NTU, depending on range | |
Accuracy | 0.001-40NTU: ±2% of reading or ±0.015NTU, whichever is greater; 40-100NTU: ±5% of reading | |
Temperature | Principle | Thermistor Method |
Range | 0℃~80℃ | |
Resolution | 0.1℃ | |
Accuracy | ±0.5℃ |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 西安东升 |
| Spec: | DS-WQ2000 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |