DS-CR6+200 Hexavalent Chromium Water Quality Online Automatic Monitor
Contact Info
- Add:西安市高新区锦业二路逸翠尚府北区8号楼1单元1701, Zip: 710000
- Contact: 魏经理
- Tel:15686405979
- Email:1528126689@qq.com
Other Products
MeasurementPrinciplePrinciple
Main FeaturesFeatures
uUses a sapphire valve core ten-way valve with high wear resistance, resistance to strong acids and alkalis, organic solvents, long service life, and high stability;
uEmploys a specially designed reference method, where minor displacement or contamination of the digestion chamber does not affect measurement values;
uFeatures waste liquid separation function, enabling separate discharge of high-concentration reaction waste and low-concentration cleaning waste;
uIncludes dynamic dilution function, automatically adjusting dilution ratio based on the on-site set range;
uOffers quality control functions such as built-in spiked recovery, blank testing, standard sample verification, zero point verification, span verification, parallel sample testing, and supports automatic, manual, and remote initiation of quality control;
uProvides automatic fault diagnosis and abnormal information upload and recording functions, including component failure, over-range alarm, exceedance alarm, reagent shortage alarm, etc.;
uIncludes overheating protection: automatically cuts off power to the heater if the temperature is too high;
uAfter abnormal reset or power restoration, the instrument automatically discharges residual reactants and returns to the initial working state. Features power-on alarm function, generating an alarm message upon restart after power loss;
uRecords process logs, measurement data, and alarm information, storing data for over 3 years with hourly recordings, and supports fast data export via USB drive;
uHas three-level management permissions;
Typical ApplicationsApplication
uWidely used for online monitoring of Cr6+ concentration in drinking water, surface water, municipal sewage inlets and outlets, pollution source discharge points, and industrial wastewater outlets.
Technical SpecificationsSpecification
Range | 0-5mg/L (customizable) |
Indication Error | ≤±10% |
Quantitation Limit | ≤0.01mg/L |
Repeatability | ≤2% |
Resolution | 0.0001mg/L (decimal places 2-4 configurable) |
Display | 7-inch touch screen |
Measurement Modes | Continuous mode, periodic mode (1h~30d adjustable), fixed-point mode, controlled mode, manual mode optional |
Calibration Modes | Manual calibration and automatic calibration |
Maintenance Cycle | Generally once per month |
Power Supply | 220V±22V, 50Hz±0.5Hz |
Output | 4-20mA output |
Communication Methods | MODBUS RS-232, RS-485 |
Environmental Conditions | Indoor with adjustable temperature, recommended temperature+5~40℃; humidity≤85% (non-condensing) |