Features
lEquipped with a sapphire valve core ten-way valve, featuring high wear resistance, resistance to strong acids and alkalis, organic solvents, long service life, and high stability;
lUtilizes a specially designed reference method, where minor displacement or contamination of the digestion chamber does not affect measurement values;
lIncludes waste separation functionality, enabling separate discharge of high-concentration reaction waste and low-concentration cleaning waste;
lFeatures dynamic dilution capability, automatically adjusting dilution ratios based on on-site range settings;
lIncorporates quality control functions such as built-in spiked recovery, blank testing, standard sample verification, zero-point verification, span verification, and parallel sample testing, with options for automatic, manual, or remote initiation of quality control;
<lSupports automatic fault diagnosis and upload/recording of abnormal information, including component failures, over-range alarms, exceedance alarms, reagent shortage alarms, etc.;
uIncludes overheating protection: automatic power cutoff for the heater when temperature is too high;
uAutomatic discharge of residual reactants and restoration to initial working state after abnormal reset or power restoration. Features power-on alarm functionality, generating an alarm message upon reboot after power loss;
uProvides process logs, measurement data recording, and alarm information storage. With data recorded hourly, storage capacity exceeds 3 years, and data can be quickly exported via USB drive;
lIncludes three-level management permissions;
Typical Applications
lOnline monitoring of total nitrogen concentration in drinking water and surface water;
lOnline monitoring of total nitrogen concentration at municipal wastewater inlets/outlets, pollution source discharge points, and industrial wastewater outlets.
Technical Parameters
Range | 0-500mg/L (customizable) |
Indication Error | ≤±5% |
Repeatability | ≤2% |
Resolution | 0.0001 mg/L (decimal places configurable from 2 to 4 digits) |
Display | 7-inch touch screen |
Measurement Modes | Continuous mode, periodic mode (adjustable from 1h~30d), fixed-point mode, controlled mode, manual mode optional |
Calibration Modes | Manual calibration and automatic calibration modes |
Maintenance Cycle | Typically once per month |
Power Supply | 220VAC±10%, 50HZ±10% |
Output | 4-20mA output |
Communication Methods | MODBUS RS-232,RS-485 |
Environmental Conditions | Indoor environment with adjustable temperature, recommended temperature +5~40℃; humidity ≤85% (non-condensing) |
Dimensions | 520*390*1475mm (L*W*H) |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 西安东升 |
| Spec: | DS-TN200 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |