Automatic Colloidal Layer Index Tester
Contact Info
- Add:东杨工业区瓯江路222号, Zip:
- Contact: 郑磊
- Tel:13603925078
- Email:15232611@qq.com
Other Products
Technical Features:
Complies with the technical requirements of GB/T 479-2000 "Method for Determination of Caking Index of Bituminous Coal";
Uses Siemens PLC as the core of the central control system, ensuring high reliability; the entire process requires no manual intervention, enabling automatic measurement and control of test temperature, X-displacement curve, and plastic layer thickness (Y-value), with real-time display of the overall equipment operation status on the computer screen;
Dynamic force robotic arm simulates the manual measurement process, enabling accurate measurement of lean coal and meager lean coal with Y-values less than 7;
Rapid replacement of furnace bricks and automatic determination of test end time improve work efficiency;
Horizontal arrangement of coal cups and vertical lever pressure application facilitate convenient coal loading, completely eliminating graphic distortion caused by cup jamming and inaccurate measurement data;
Mechanical loading and unloading of weights reduce the operator's labor intensity;
The robotic arm uses a ball screw mechanism (other companies use simple screw mechanisms with short service life);
Automatically saves temperature and volume dynamic data from each test for query at any time, allowing recovery of test data and curve records, and printing of result reports;
Features ultra-low temperature fault protection, ultra-high temperature fault protection, and thermocouple fault protection functions;
Includes temperature and displacement calibration functions.
Technical Parameters:
Temperature Control Section:
Temperature measurement resolution: 0.1°C;
Temperature measurement accuracy: Grade 0.5;
Temperature control process: Room temperature ~250°C, 30min, heating rate: approx. 8°C/min; test time resets and restarts timing, 250~730°C, 160min, heating rate: 3°C/min;
Temperature control accuracy: ±1°C;
X-Displacement Indicators:
Measurement range: 0~100mm, imported cable extension position sensor;
Measurement accuracy: 0.001.FS;
Sensor type: Photoelectric digital isolation type;
Inherent resistance: ≈0.1g;
Y-Displacement Indicators:
Measurement range: 0~200mm, imported cable extension position sensor;
Measurement accuracy: 0.001.FS;
Sensor type: Photoelectric digital isolation type;
Inherent resistance: ≈0.1g;
Measurement error: ≈0;
Pressure Sensor Indicators:
Measurement range: 0-50kg
Intelligent Robotic Arm Characteristics:
Y-value error: When Y≥20mm, error <2mm
When Y<20mm, error <1mm
X-value error: △<3mm
Silicon Carbide Rod Characteristics:
Resistance: 3-8Ω;
Effective heated length: 150mm, diameter 8mm;
Cold end length: 60mm, diameter 16mm;
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 天润科技 |
| Spec: | TRJC-8A |
| Stock: | 300 |
| Manufacturer: | |
| Origin: | China / Henan / Hebishi |