China Dual-channel controller - China Supplier
China Dual-channel controller - China Supplier

Dual-channel controller

Price:2000
Industry Category: Measurement-Analysis-Instruments
Product Category:
Brand: OLDHAM
Spec: SV4B


Contact Info
  • Add:嘉兴市城东路1137号, Zip:
  • Contact: 容志芳
  • Tel:13957320530
  • Email:office@jxlead.com

Other Products

Description
Additional Information

SV4B French OLDHAM Dual-Channel Controller


SV4B French OLDHAM Dual-Channel ControllerMeasurement values from sensors and input modules are instantly managed by MX43. When the measured value reaches the set threshold, audible and visual alarms are activated. Simultaneously, one or more corresponding relays are triggered to control internal or external remedial measures preset by the user.

Programming of ZY detectors is completed via the workstation software COM43.

SV4B French OLDHAM Dual-Channel Controller Panel

Dimensions

370x299x109mm (Wall-mounted), 19", 4U ; 482.8x177x192.5 mm (Rack-mounted)

Weight

4kg (Wall-mounted), 2kg (Rack-mounted)

Protection Rating

IP55 (Wall-mounted), IP31 (Rack-mounted)

Cable (Wall-mounted)

12 M16 cable glands, 4 to 8 mm² (8 to 11 AWG) outer diameter cable

6 M20 cable glands, 6 to 12 mm² (7 to 9 AWG) outer diameter cable

Display

LCD backlit display + smart keys

Reverse video display on fault

User customizable (simultaneous display of 1 to 16 channels, fixed or scrolling, or events...)

Bar graph and alarm thresholds

Visual Indicators

7 LEDs per line

1 LED fault indicator

1 LED fault indicator

Buttons

5 smart keys, 1 audible alarm acknowledge/reset button

Operation

Operating Temperature

-20°C ~ +50°C

Storage Temperature

-20°C ~ +50°C

Humidity

5 ~ 95% RH

Power Supply

100-240 Vac 50-60 Hz or 21-28 Vdc, max 112 W

Battery

Embedded backup power option (0.6 Ah)

Power Consumption

Min 500 mA (without modules)

Measurement Lines

Digital Lines

Up to 8 lines

RS-485 communication, proprietary protocol, 9600 baud

2 twisted pairs

Analog Lines

Up to 8 lines

0-23 mA analog signal input (4 to 20 mA reserved for measurement) 120 Ohm load impedance

2 or 3 cores

Voltage (Typical)

21 ~ 28Vdc

Current Output per Line

500 mA

Total Current Output

2.4 A continuous (3.2 A intermittent)

Alarms

Per Channel

5 alarm levels (A1, A2, A3, over-range, under-range) + fault

Non-ambiguous reading option for combustible gas monitoring

Programmable Thresholds

Instant average, rising or falling alarms, manual or automatic

Onboard Relays

5 fully programmable alarm relays

1 fault relay (non-configurable)

Dry contact relays

DPCO relays, contact rating 250 Vca 2 A (inductive)

Digital Output

RS-485 Modbus RTU serial link

Certifications

EMC

Complies with EN50270

ATEX

INERIS 13 ATEX 0048

Low Voltage Directive

Complies with EN61010

CSA

In progress

SV4B French OLDHAM Dual-Channel Controller Chinese Operation Interface
Modular design, flexible configuration; supports remote installation
32 digital inputs, 16 logic inputs
24 programmable relay outputs
6 alarm levels

Industry Category Measurement-Analysis-Instruments
Product Category
Brand: OLDHAM
Spec: SV4B
Stock: 100
Manufacturer:
Origin: China / Zhejiang / Jiaxingshi
About Toocle.com - Partner Programme - Old Version
Copyright © Toocle.com. All Rights Reserved.
(浙)-经营性-2023-0192