CV-3A Preform Thickness Tester
Contact Info
- Add:济南市历城区, Zip: 250000
- Contact: 李小姐
- Tel:13335118700
- Email:3965028388@qq.com
Other Products
CV-3A Preform Thickness Tester
Product Introduction
1 Equipment Purpose
Glass containers such as ampoules, vials, and infusion bottles occupy a significant share in the pharmaceutical packaging field. Due to their excellent stability, glass containers have become a crucial component of the pharmaceutical packaging industry. The thickness of glass bottles is one of the factors affecting the stability of pharmaceutical glass containers. Therefore, measuring the thickness of glass bottles is one of the fundamental performance requirements for pharmaceutical glass containers.
The bottom and wall thickness tester is designed in accordance with the national pharmaceutical packaging material standards (2015). Its purpose is to measure the bottom thickness and wall thickness of pharmaceutical glass bottles.
2 Equipment Technical Specifications
Specification | Parameter |
Measuring Range | 0-12.7mm |
Graduation Value | 0.001mm |
Sample Diameter | 10mm-230mm (other sizes customizable) |
Sample Height | 10mm-300mm (other sizes customizable) |
Ambient Temperature | 15-50℃ |
Overall Dimensions | 420(L)mm × 250(W)mm × 450(H)mm |
Net Weight | 30kg |
Structural Composition
Mainly consists of a bottom thickness measuring head, wall thickness measuring head, measuring rod, main unit, etc.
Equipment Installation
1 Equipment Operating Environment Requirements
l Ambient temperature 23±2℃.
l Ambient humidity 50% RH.
l Avoid using on the same power line as high-power equipment.
l The environment should be clean, free from strong electromagnetic interference, and corrosive gases.
l The test platform should be stable, free from any vibration, and away from heat sources.
2 Equipment Installation Process
l Carefully open the packaging box cover and check the technical documents and accessories against the packing list.
l Remove the instrument and place it on a stable and firm work surface, ensuring there is at least 0.5 meters of operating space around it. Keep the instrument level and secure.
l Read and thoroughly understand the user manual. Operators must carefully read through this manual to understand the installation process and master all operational requirements.
l Properly store the packaging box for future use in return shipping for maintenance.
Label: CV-3A Preform Thickness Tester
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 济南兰博 |
| Spec: | CV-3A |
| Stock: | 10 |
| Manufacturer: | |
| Origin: | China / Shandong / Jinanshi |