RGYC-1 Direct Reading Moisture Content Analyzer for Boiler Flue Gas
Contact Info
- Add:仙山东路30号, Zip: 266000
- Contact: 安超
- Tel:13165062660
- Email:67168224@qq.com
Other Products
RGYC-1 Moisture Content Tester (Gravimetric Method)
The RGYC-1 Moisture Content Tester uses the gravimetric method to determine the moisture content in flue gas from stationary pollution sources. This tester primarily consists of two parts: a flue gas moisture sampling probe and a gravimetric weighing module for moisture content. The front end of the sampling probe is heated, while the rear end is cooled at the desiccant absorption tube to ensure moisture in the flue gas is absorbed by the tube. The weighing module measures the weight of the absorption tube before and after sampling, wirelessly receives sampling data from the probe, and automatically calculates moisture content based on the sampling and weight data.
Relevant Standards
HJ/836-2017 "Determination of Low Concentration Particulate Matter in Exhaust Gas from Stationary Pollution Sources—Gravimetric Method"
GB16157-1996 "Determination of Particulate Matter in Exhaust Gas from Stationary Pollution Sources and Sampling Methods for Gaseous Pollutants"
HJ/T47-1999 "Technical Specifications for Flue Gas Samplers"
Product Features
Simple on-site operation, with data available in 5 to 10 minutes.
National standard method—gravimetric measurement of moisture content is accurate and reliable, suitable for calibrating sensor-based moisture testers.
Wide adaptability to working conditions, including high-temperature and complex environments.
The sampling probe features heating to ensure moisture passes through the flue in vapor form and is absorbed by the desiccant tube.
Wireless transmission function—sampling data is wirelessly transmitted to the moisture tester, and moisture content data is wirelessly sent to a printer.
The moisture tester calculates the net weight of moisture based on the weight difference of the absorption bottle before and after sampling, and automatically computes moisture content from the sampling data.
Titanium filter element removes particulate matter from the flue gas.
Portable power supply box available as an optional accessory.
Main Parameters | Parameter Range | Main Parameters | Parameter Range |
Moisture Content Range | 0~100 | Moisture Resolution | 0.1% |
Sampling Time | 0~24h | Desiccant | Silica Gel Particles |
Flow Range | (0.5~2) NL/min | Power Supply | AC 220 V±10% |
Weighing Resolution | 0.001g | Weighing Accuracy | 0.01g |
Probe Heating Temperature | 80-180℃ adjustable | Operating Temperature Range | 0~200℃ |
Qingdao Rongguang Electronic Technology Co., Ltd. provides technical support and comprehensive after-sales service for this instrument!
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 青岛容广13165062660 |
| Spec: | RGYC-1 |
| Stock: | 9999 |
| Manufacturer: | |
| Origin: | China / Shandong / Qingdaoshi |