CoMetro6000 Post-Column Derivatization System
Other Products
CoMetro 6000 Series Post-Column Derivatization System (Post-Column Derivatization Unit) Features:
Ø Quick and easy setup
Connect one inlet from the column and one outlet to the detector, and installation is complete.
Ø Modular design
Optional single or dual reaction cores. Capable of performing two-step reaction processes simultaneously, including heated reactions and room-temperature reactions.
Optional one, two, or multiple reagent pumps.
Ø Reduced number of fittings and tubing connections to prevent leaks.
Ø All connection ports are located on the front panel for user convenience.
Ø Front panel display for easy setup
The reactor front panel displays the set temperature and actual temperature of each reaction core.
The reagent pump front panel displays the flow rate, pressure, and pressure limit for each reagent.
Ø External control
Pump flow rates and reactor temperatures can be controlled via the front panel or externally through RS232.
Ø Modular reactor, requiring only two connections to replace the reactor core.
Ø Designed reaction core reduces peak broadening.
Ø Full PEEK flow path system.
CoMetro 6000 Series Post-Column Derivatization System (Post-Column Derivatization Unit) Reactor Features:
Continuous reaction loop, sealed Multi-directional flow for mixing
Front panel digital control: Temperature setting Temperature display Ready indicator
Heating Core | |
Reactor Volume | 0.15/0.5/1.5ml (optional) |
Reaction Volume Accuracy | ±5% |
Temperature Control Range | Ambient +10°C to 150°C |
Temperature Repeatability | ±0.4°C |
Temperature Accuracy | ±1°C |
Safety Temperature Cutoff | 160°C |
Stabilization Time | Ready indicator displayed within 30 minutes at 150°C |
Input | Front panel or RS232 |
Reagent Pump Features and Parameters:
Automatic online self-cleaning to extend seal life
Built-in degassing valve design (PEEK)
Built-in pressure sensor
Stepper motor drive, electronic fast priming
Digital keypad display: Flow rate setting, pressure setting, set pressure upper/lower limits
Flow Rate | 0.001-10 ml/min |
Pressure | 0-4000 PSI |
Flow Path | Full PEEK flow path system |
Flow Precision | 0.2% RSD |
Flow Accuracy | ±1% |
Input | Remote control and monitoring via RS232 interface |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | |
| Spec: | |
| Stock: | |
| Origin: |