HCR-NY002 Pesticide Oxidation/Reduction Chemical Incompatibility Tester
Other Products
HCR-NY002Pesticide Oxidation/Reduction Chemical Incompatibility Tester
This instrument is designed and manufactured in accordance with the standards of NY/T 1860.10-2016 Guidelines for the Determination of Physical and Chemical Properties of Pesticides.
The testing method of this instrument is a qualitative evaluation method. It determines the oxidation-reduction/chemical incompatibility of pesticides by observing whether hazardous reactions occur when the tested pesticide comes into contact with common oxidizers, reducers, fire extinguishing agents, and solvents. The mass ratio between the test substance and the selected reaction reagent should be sufficiently large (generally 3:1 to 1:1) to simulate the worst-case scenario, and the amount of the test substance should meet the requirements for temperature measurement.
This instrument is compact in size, easy to operate, and highly accurate, making it very suitable for laboratory use.
Performance and Features:
1. This instrument has an integrated structure and can test six samples simultaneously or individually.
2. It uses a digital display high-precision sensor to measure temperature, ensuring high accuracy and intuitive display.
3. It employs an electronic flow sensor to detect changes in gas flow, allowing observation of instantaneous and total gas flow through the LCD screen.
4. It uses standard ground glass test tube interfaces, ensuring tight connections, convenience, and ease of operation.
5. This instrument features a simple structure, aesthetically pleasing appearance, reasonable design, easy operation, and high safety and reliability.
Technical Parameters
Implementation Standard NY/T 1860.10-2016
Operating System WIN7 System Chinese Software Menu-Based List Operation
Detection Method: Touchscreen computer automatically controls temperature and gas detection
Temperature Detection High-Precision Temperature Sensor
Testing Accuracy: ±0.1°C, ±0.1ml
Testing Units: Six Tubes
Gas Detection: Flow sensor for real-time gas release detection and cumulative gas volume recording
Power Supply: AC220V±10% 50Hz
Power Consumption: 100W
Working Environment: Laboratory with ventilation and safety protection
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | |
| Spec: | |
| Stock: | |
| Origin: | China / Henan / Zhengzhoushi |