High Purity Water Treatment Instrument RS
Contact Info
- Add:山东省枣庄市滕州市东沙河街道东沙河村首康康路10号, Zip: 277500
- Contact: 李书典
- Tel:0632-5251855
- Email:sdxinruiq@163.com
Other Products
The laboratory high-purity water purifier is a water purification device with higher purity than ordinary water purifiers, primarily used for the final rinsing of laboratory glassware, preparation of buffer solutions and chemical reagents, preparation of microbial culture media, pure water for hydrogen generators, indoor humidifiers, high-pressure sterilizers, and drinking water for laboratory animals.
Detailed Introduction
The NewRay laboratory high-purity water purifier equipment features a microcomputer controller that allows manual and automatic switching operation. Its control functions include: automatic shutdown in cases of no water or full water, low-pressure and over-pressure automatic protection, and customizable water output indicators. If the set value is not met, the water is recirculated through a backpressure valve for further treatment. Once the set requirements are met, the water collection valve opens for use. The reverse osmosis (RO) elements, high-purity precision polishing resin, main components, machines, pumps, and instruments of this equipment can all be selected as imported parts, ensuring reliable quality.
The NewRay laboratory high-purity water purifier is designed with different pre-treatment functions, equipped with various filtration units tailored to different water qualities, such as softening and strong activated carbon adsorption units for high hardness and high turbidity water.
Small microcomputer control function; operated by a small microcomputer, the machine automatically shuts down in cases of water shortage or full water, and runs an automatic flushing function each time it starts.
Compared to traditional electric thermal distillers, it produces water with higher purity and stability; it is easy to operate and does not require dedicated supervision; it has no electric heating tubes, no scaling issues, and offers good safety.
Features of the NewRay laboratory high-purity water purifier system:
· Uses a silent pump, ensuring no vibration, noise, or electromagnetic radiation during operation, with no impact on the working environment.
· Employs reverse osmosis membranes with stable performance, high desalination rate, and long service life.
· Uses exchange resin to produce (ultra) purification columns with high exchange capacity and superior water quality.
· Features an all-metal chassis with electrostatic spray painting, effectively preventing rust from water vapor.
· Separates water and electrical circuits, avoiding the impact of humid environments on the electrical system.
Performance Features of the RS Series High-Purity Water Purifier
Equipped with power indicator, pump operation indicator, low-pressure indicator, and system flushing indicator functions.
2. Uses imported ion filtration membranes and high-pressure, silent pure water pumps for more reliable operation and broader applicability to various water qualities;
3. Features a system flushing function: before producing water each time it starts, it automatically flushes for 18 seconds before beginning water production;
4. Includes a system protection function: automatically shuts down when water pressure is too low or there is no water, and automatically restarts when water supply resumes;
5. In addition to the above features, the RS-II high-purity water purifier also has an online monitoring function, allowing real-time detection of the output water quality, displaying water quality (conductivity online monitoring, with higher accuracy) and temperature on an LCD background.
RS Series High-Purity Water Purifier (Using Tap Water as Source)
Name High-Purity Water Purifier/High-Purity Water Generator
Product Model RS-Series (10-100L/H)
Inlet Water Requirements Municipal tap water; TDS<200ppm; Water pressure>0.1Mpa; Water temperature: 5-45℃
Pure Water Quality Conductivity <1μs/cm (25℃)
pH Value (25℃) Requirement: None; Test Result: 7.3
Oxidizable Substances (as O) Technical Requirement: ≦0.08mg/L; Test Result: ≦0.08mg/L
Residue on Evaporation (105±2℃) Technical Requirement: ≦1.0mg/L; Test Result: 0.7mg/L
Test Standard: GB/T6682-2008,
Technical Specifications Power/Power Consumption 25-135W
Water Production Capacity 10-100L/H
Dimensions/Weight 50*45*78 (cm) /52KG
Control System Operation Mode Display Power, water shortage, flushing, etc.;
Safety Performance Automatically enters standby mode when there is no water or low water pressure;
Purification System Pre-treatment Unit External softening filter (recommended for areas with high raw water hardness); 1μm deep PP filter element*2 + precision activated carbon filter element
RO Reverse Osmosis Unit Original US-made RO reverse osmosis membrane/membrane module ;
Post-Purification Unit High-purity column (reusable after regeneration treatment)
Standard Configuration Main unit + accessory kit
Note: 1. Inlet water quality affects output water quality and the service life of the filter column; 2. When the source water quality is poor (TDS greater than 200ppm), the RO pure water quality may drop to above 10μs/cm²; 3. When the raw water TDS>500, it is recommended to use a water softener;
Features of the High-Purity Water Purifier
Power Consumption Low energy consumption: 220V/40W
Water Purity 0.6μs/cm (25℃)
Water Quality Standard (GB6682-2008) Exceeds laboratory Grade 2 water requirements
Wastewater Ratio 1:3 (1 represents water production capacity)
Water Stability High purity and strong stability of output water
Operability Easy to operate, no need for dedicated supervision
Safety No electric heating tubes, no scaling issues, good safety
| Industry Category | Environment |
|---|---|
| Product Category | |
| Brand: | 山东新瑞 |
| Spec: | RS |
| Stock: | 1000 |
| Manufacturer: | |
| Origin: | China / Shandong / Zaozhuangshi |