China Instrument for testing sulfur content in fluorite powder - China Supplier
China Instrument for testing sulfur content in fluorite powder - China Supplier China Instrument for testing sulfur content in fluorite powder - China Supplier China Instrument for testing sulfur content in fluorite powder - China Supplier China Instrument for testing sulfur content in fluorite powder - China Supplier China Instrument for testing sulfur content in fluorite powder - China Supplier

Instrument for testing sulfur content in fluorite powder

Price:Negotiable
Industry Category: Measurement-Analysis-Instruments
Product Category:
Brand: 英特仪器
Spec: SY-9A


Contact Info
  • Add:鹤壁市长风路, Zip: 458000
  • Contact: 师文亮
  • Tel:0392-2700039
  • Email:598945101@qq.com

Other Products

Description
Additional Information

Fluorite Powder Sulfur Content Testing Instrument Fluorite Powder Sulfur Content Testing Instrument Hebi Yinte Instrument & Meter Factory Fluorite Sulfur Content Analyzer Fluorite Sulfur Content Tester

The correct operation is to first turn on the main unit, then open the computer program. If the computer program is not connected to the main unit, some operations cannot be performed, as shown below:

Under the Options menu, there are 4 submenus: Name Management, Parameter Settings, System Calibration, and Exit Program.

Name Management: Allows adding, deleting, and modifying user names, analyst names, and sample names.

Under the Parameter Settings menu, there are Furnace Temperature Settings and Serial Port Settings menus.

Furnace Temperature Settings are used to set the temperature of the pyrolysis furnace in the sulfur analyzer.

The System Calibration menu is used to adjust the accuracy of the sulfur analyzer. It includes Overall Calibration and Segmented Calibration menus.

Overall Coefficient Calibration is used when test results (high sulfur, low sulfur) are consistently high or low.
Segmented Calibration is used when sulfur values in a specific range are high or low. For specific settings, refer to "Common Issues and Maintenance Methods of Sulfur Analyzers."

If the instrument is connected to an external balance, sample weight can be automatically input. Right-click in the balance data window and select the "Add Sample Weight" menu to automatically and continuously input sample weight information, as shown below:

After entering the sample information, click "Start Test." The instrument will then automatically complete the required operations.

Industry Category Measurement-Analysis-Instruments
Product Category
Brand: 英特仪器
Spec: SY-9A
Stock: 80
Manufacturer:
Origin: China / Henan / Hebishi
About Toocle.com - Partner Programme - Old Version
Copyright © Toocle.com. All Rights Reserved.
(浙)-经营性-2023-0192