DBZX-520 Online Hydrogen Content Detection Device for Inner Cooling Water Tank
Other Products
Product Introduction
DBZX-520The online hydrogen content detection device for internal cooling water tanks is specifically designed for the unique characteristics of hydrogen-cooled generator units' internal cooling water tanks (hydrogen content range 0 to 10% vol). It is an online detection device developed using the thermal conductivity principle to measure hydrogen concentration and hydrogen leakage in internal cooling water tanks. This device not only detects the real-time hydrogen content in the internal cooling water tank but also measures the daily hydrogen leakage and total cumulative flow. Equipped with an embedded microprocessing system, this product is powerful and ideal for detecting hydrogen leakage in internal cooling water tanks.
Features
◆ The instrument uses an advanced constant-temperature thermal conductivity cell with temperature compensation, ensuring excellent stability.
◆ Unique large true-color touch LCD screen with Chinese display for convenient operation.
◆ The field transmitter adopts an explosion-proof design to ensure safety.
◆ On-site calibration is available to ensure reliable instrument performance.
◆ Automatic data saving with historical data traceability function, capable of storing over 10,000 records.
◆ The field internal cooling water content and flow can be installed either separately or as an integrated unit for detection.
◆ Unique free input function for cumulative time, meeting special measurement needs.
◆ Capable of detecting hydrogen leakage, daily hydrogen leakage, cumulative hydrogen leakage, data storage, and remote transmission.
Technical Parameters
◆ Measurement range: H2 content in air 0 to 20% (customizable range).
◆ Measurement accuracy: ±0.1%.
◆ Resolution: 0.01%.
◆ Flow measurement range: 0.016 to 2.5 m³/h.
◆ Signal output: 4-20 mA or RS-485.
◆ Alarm output: Relay output.
◆ Power supply: 220 VAC ±10% or DC 24V ±10%.
◆ Explosion-proof rating: Exia II C T4 Ga.
◆ Remote gas flow meter (optional).
◆ Nominal flow: 1.6 (Qn m³/h).
◆ Maximum flow: 2.5 (Qmax m³/h).
◆ Minimum flow: 0.016 (Qmin m³/h).
◆ Measurement error: Class B ±1.5%.
| Industry Category | Electrical-Equipment-Supplies |
|---|---|
| Product Category | |
| Brand: | 迪邦 |
| Spec: | |
| Stock: | |
| Origin: | China / Henan / Zhengzhoushi |