SIP Online Sterilization Temperature Validation Clamp Clean Pipeline Reaction
Contact Info
- Add:安徽省合肥市肥西县经济开发区繁华大道立恒工业广场二期B1栋3楼, Zip: 230000
- Contact: 汪女士
- Tel:18456533083
- Email:18456533083@163.com
Other Products
Online Sterilization:
· One-step sterilization (suitable for small systems);
· Sterilization in multiple steps, such as: concentrated preparation, diluted preparation, filling, mixing tanks, storage tanks, sterilizing filters; filling machines and multi-pipeline systems, etc.;
· Overlapping areas of sterilization;
Key Points of SIP Validation:
· System air evacuation to ensure all parts of the system are saturated with steam;
· Drain condensate through small pipelines/hoses (e.g., filling nozzles);
· Placement of biological indicators at appropriate locations (e.g., filter screens in pipelines located at the front end of steam valves);
· Sterilization of filters (diaphragm pressure difference);
· Drying of filters after sterilization;
· Non-contaminating integrity testing;
Validation Content for Online Sterilization (SIP):
· Verification of distribution across all parts of the system
· Temperature validation: Sterilization temperature testing
· BI testing: Biological indicator challenge testing
· Pressure validation: Online sterilization pressure testing
· F0 value/kill rate: Microbial lethality rate
· Sterilization time: Validation of effective sterilization time
· Saturated steam conditions (temperature and pressure)
· Sterilization cycle (duration maintained)
· System drying
· Inactivation of all biological indicators
· Integrity of filters post-sterilization
· Pressure increase during cooling with sterile air
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | |
| Spec: | CVT-SIP-25 |
| Stock: | 99 |
| Manufacturer: | |
| Origin: | China / Anhui / Hefeishi |