XYZP-IB Computerized Medium Frequency Therapy Apparatus
Contact Info
- Add:北京市海淀区紫雀路33号院3号楼四层, Zip:
- Contact: 陈贞
- Tel:13258102986
- Email:465364378@qq.com
Other Products
I: XYZP-IB Computer Intermediate Frequency Therapy Apparatus Product Performance
1. Operating Conditions:
a) Ambient temperature range: +5℃ to +40℃; b) Ambient humidity range: ≤80%;
c) Atmospheric pressure range: 700hPa to 1060hPa;
d) Rated voltage: a.c.220V, rated frequency: 50Hz.
2. Rated input power: 60VA.
3. Main unit dimensions: length 280mm, width 205mm, height 80mm, tolerance ±20mm.
4. Electrode wire: length 1300mm, tolerance ±50mm.
5. Square electrode plates:
XX-107*72: length 107mm, width 72mm, tolerance ±10%; XX-85*55: length 85mm, width 55mm, tolerance ±10%.
6. Operating frequency: 2kHz to 10kHz, single frequency tolerance ±10%.
7. Modulation frequency: 0 to 150Hz, single frequency tolerance ±10% or ±1Hz, whichever is greater.
8. Intermediate frequency carrier waveform: bidirectional square wave, pulse width 50us to 250us, tolerance ±10%; modulation waveforms include sine wave, square wave, triangular wave, exponential wave, sawtooth wave, spike wave, and continuous wave.
9. Modulation modes: continuous, intermittent, burst, frequency conversion, dense-sparse, and alternating modulation. 10. Intermediate frequency modulation depth: 0%, 25%, 50%, 75%, 100%, tolerance ±5%. 11. Intermediate frequency therapy prescriptions: 35 fixed prescriptions.
12. Intermediate frequency output current: under a 500Ω load, each channel output current not exceeding 100mA, output intensity adjustable in 0 to 99 levels.
13. Output current stability: the output current variation rate under different loads should not exceed 10%.
14. Intermediate frequency output peak voltage: when measured under open circuit conditions, the intermediate frequency output peak voltage must not exceed 500V. 15. Operation: when output is set to maximum, operate open circuit for 10 minutes and then short circuit for 5 minutes, the therapy apparatus should function normally.
16. Electrode plate temperature: 38℃ to 55℃, adjustable in 6 levels, tolerance ±3℃.
17. Electrode plates: should be selected as qualified products with Class I medical device filing certificates.
18. Treatment time: intermediate frequency treatment time is preset in the prescription, with an audio prompt at the end and output cessation, tolerance ±1min.
19. Load impedance: the effective load impedance of this therapy apparatus is 500Ω, and the impact on technical indicators is negligible when the rated load impedance varies within ±10%.
20. Operating mode: continuous operation.
21. Continuous operating time: verified through actual operation, should be no less than 4 hours.
| Industry Category | Furniture |
|---|---|
| Product Category | |
| Brand: | 翔宇 |
| Spec: | 13258102986(微信同号) |
| Stock: | 1000 |
| Manufacturer: | |
| Origin: | China / Henan / Anyangshi |