Single-Use Thoracic Drainage Device Compliance Tester TDE-02 testing instrument is developed in accordance with the requirements of Appendix C in the standard YY0583-2005 "Single-Use Thoracic Drainage Device - Water-Seal Type".
Thoracic Drainage Device Compliance Tester
A single-use water-seal thoracic drainage device is a medical instrument primarily used for patients with pneumothorax, pleural effusion, or those requiring closed drainage after surgery. It consists of an overall water-seal bottle structure, including an effusion collection bottle, water-seal bottle, pressure regulation bottle, ear hooks, support feet, and drainage tubing. This device is typically used in conjunction with thoracic drainage catheters to help drain fluid or gas from the pleural cavity. Indications for the single-use water-seal thoracic drainage device include traumatic hemopneumothorax, spontaneous pneumothorax, and pleural effusions due to various causes. The device comes in multiple specifications, including Type I double-tube, Type I single-tube, Type II double-tube, and Type II single-tube.
YY/T0583.1-2015 specifies the testing requirements for water-seal single-use thoracic drainage devices. The compliance tester refers to Appendix C of YY/T0583.1-2015.
What testing instrument is used for the compliance test of single-use thoracic drainage devices?
The Single-Use Thoracic Drainage Device Compliance Tester TDE-02, developed and produced by Zhongce Electromechanical, is used to measure the compliance performance of thoracic drainage devices. This testing machine also supports error measurement, retention, opening pressure, release speed, gas leakage, and other test items.
Main Parameters of Single-Use Thoracic Drainage Device Compliance Tester TDE-02
Product Name: Water-Seal Valve Thoracic Drainage Tester
Pressure Range: -100 to 10 kPa (standard)
Pressure Accuracy: 0.5% FS
Flow Range: 0 to 1000 mL/min (standard)
Flow Accuracy: 0.5% FS
Air Source Pressure: ≤ 0.7 MPa (air source to be provided by the user)
Air Source Interface: Φ6 mm polyurethane tube
Dimensions: 390 mm (L) × 410 mm (W) × 190 mm (H)
Power Supply: As indicated on the product nameplate
Testing Principle: First, measure the total internal volume of the system in its ready-to-use state. Then, according to Appendix D, measure any leaks by blocking all outlets, extracting air from the system to reach the specified pressure, and recording the suction volume to determine the compliance of the drainage device.
Product Configuration
Standard Configuration: Main unit, micro printer, air source connector
Note: The air source interface of this machine is an Ф8 mm polyurethane tube; the air source must be provided by the user.
Detailed Information on the Single-Use Thoracic Drainage Device Compliance Tester:http://www.pubtester17.com/pro/579.html
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 众测机电 |
| Spec: | TDE-02 |
| Stock: | |
| Origin: | China / Shandong / Jinanshi |