Y09-110 Laser Particle Counter
Contact Info
- Add:苏州市吴江经济开发区江陵街道联兴村渡船桥花苑东区23号, Zip: 215216
- Contact: 高新
- Tel:13776141586
- Email:2319039989@qq.com
Other Products
Y09-110 Laser Particle Counter
Product Overview
- Light Source: Semiconductor Laser
- Sampling Flow Rate: 28.3L/min
- Particle Size Channels: Eight channels for simultaneous counting
- Display: Large LCD touch screen (LCD)
- Power Supply: Rechargeable lithium battery 16.8V
- Data Storage: Can store at least 100,000 sets of data
Product Features
This instrument is a successful product developed in collaboration with domestic universities and research institutions in accordance with the ISO 14644-1:2015 international standard. The entire product features a housing made of 316L stainless steel. The laser sensor uses a full semiconductor laser and semiconductor photodiode receiver, ensuring the stability of the light source and the accuracy of the signal receiver. It significantly reduces stray scattered light within the scattering chamber, improving the signal-to-noise ratio of the sensor. The LCD touch screen control allows for clear data display and convenient operation. The sampling pump is imported from the United States, and it includes a built-in differential pressure sensor for closed-loop flow control, ensuring the flow rate remains stable within the set range. The instrument can automatically determine various purification levels based on the standards selected by the customer.
Application Range
This product is widely used in medical, pharmaceutical, electronic, chemical, optical, food, cosmetics, biological products, aerospace industries, and various testing institutions for dust particle detection.
Light Source | Semiconductor laser, lifespan greater than 10 years >30,000h |
Sampling Flow Rate | 28.3L/min (1ft3) ±5%, constant flow rate adjustable based on the resistance of the sampling tube length |
Display Method | 155×85mm high-resolution LCD touch screen display (LCD), Chinese interface, real-time display, previous cycle display, real-time concentration display, can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling times, status and other parameters, 95% UCL calculation, can directly display particle concentration (particles/m³), three-level permission management, audit trail function. |
Power Supply/Operating Power | Maximum power consumption : 20W built-in long-life lithium battery: |
Operating Time | 8 hours |
Counting Mode | Cumulative value, difference value, concentration value |
Testing Method | Single, repeat, continuous, calculate, remote, |
Unit Conversion | <极速赛车开奖直播记录官网平台,提供最精准的赛车开奖数据,让您随时掌握最新开奖信息。我们致力于为用户提供最可靠的赛车开奖服务,确保每一期开奖结果的准确性和公正性。p>Units can be converted to m³ |
Particle Size Channels | 0.1, 0.15, 0.2, 0.25, 0.3, 0.5, 1.0, 5.0μm eight particle size channels for simultaneous counting. Particle size can be customized according to customer needs. |
Sampling Cycle | 1~600
(S) Delay counting: 0~99 (S)
Self-cleaning time: ≤10 (min) |
Operating Environment | Temperature: 5~35℃ Relative humidity: 20~90%RH, no condensation.
Atmospheric pressure: 86-106KPa |
Temperature and Humidity: | Optional
(1) Temperature: -15~50℃±1℃.
(2) Humidity: 0~98%RH±5% |
UCL Settings | Sampling points (A): 2~9 points settable. Sampling times per point: (L) 2~9 times settable
Measurement position: 0-999 |
UCL Report | Complies with ISO14644-1, GB50073-2001, GMP(2010), FS209E |
Data Storage | Can store 100,000 sets of data (circular buffer) (including particle size, data, environmental data, year, month, day, time, sampling volume, data location), data is not lost after power off. |
Communication Interface | RS232, 9600 baud rate, USB interface |
Alarm Settings | The instrument has a level alarm function, can alarm for超标 after exceeding the standards of clean rooms 100, 1000, 10000, 300000, 1000000 levels,
同时符合新版GMP标准的A/B/C/D四个级别静态和动态标准的测量四种标准切换测量 |
Zero Count | Complies with JJF-1190-2008 Dust Particle Counter Metrology Calibration Specification and ISO/FDIS21501-4:2018, less than 1 per five minutes |
Coincidence Error | 5 %, at 2,000,000 particles/cubic foot |
Printing Function | Thermal printer |
Weight | 7.6kg |
Calibration | Can meet JJF-1190-2008 Dust Particle Counter Metrology Calibration Specification, GB/T6167-2007 Dust Particle Counter Performance Test Method and ISO/FDIS21501-4:2018 (E) Calibration Method for Light Scattering Dust Particle Counters in Cleanrooms, traceable to Shanghai Institute of Measurement and Testing Technology, can also be calibrated by itself or by a third-party national metrology institution, and can also be traced to the National Institute of Standards and Technology (NIST). Our company has passed the national metrology establishment standard assessment, and the company's standard dust particle counter has also participated in the national dust particle counter value comparison involving various provincial metrology institutions, and is used as a standard precision particle counter by relevant metrology and testing institutions when performing dust particle counter metrology calibration work. |
Standard Configuration | Professional isokinetic sampling head, sampling stand with sampling tube, built-in printer, aluminum alloy case, operation manual, factory test report, zero count filter, charger, sampling stand, printing paper |
Dimensions | 230*230*230(mm) (excluding air nozzle, handle and other accessories) |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 星源洁净 |
| Spec: | Y09-110 |
| Stock: | |
| Origin: | China / Jiangsu / Suzhoushi |