OPC-3016T Laser Particle Counter
Contact Info
- Add:苏州市吴江经济开发区江陵街道联兴村渡船桥花苑东区23号, Zip: 215216
- Contact: 高新
- Tel:13776141586
- Email:2319039989@qq.com
Other Products
OPC-3016T Laser Particle Counter
1. Light Source: Semiconductor Laser
2. Sampling Flow Rate: 2.83 L/min ± 5%,
Built-in flow sensor, automatic closed-loop flow control
3. Particle Size Channels: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0 μm
4. Display: 4.3-inch Capacitive Touch Screen
5. Power Supply: Rechargeable Lithium Battery 7.4V 6.4A
6. Uses imported DC brushless fan, PWM regulation control
7. Data Storage: Can store 30,000 sets of data
8. Printing Function: Thermal Printer
Product Features
This instrument is a successful product jointly developed with Nanjing University of Science and Technology and other units in accordance with the requirements of the new version of GMP (2010 edition). The entire product uses a 316L stainless steel shell. The laser sensor uses a full semiconductor laser and semiconductor photodiode receiver to ensure the stability of the light source
and the accuracy of the signal receiver. It greatly reduces stray scattered light in the scattering chamber and improves the signal-to-noise ratio of the sensor. It uses an LCD liquid
crystal touch screen control, making the display data clear and operation convenient. The instrument can automatically determine various purification levels according to customer-selected standards.
Application Range
This product is widely used in medical, pharmaceutical, electronic, chemical, optical, food, cosmetics, biological products, aerospace, and various testing institutions for dust particle detection.
Main Technical Parameters:
Light Source: Semiconductor Laser, lifespan greater than 10 years>30,000h
Sampling Flow Rate2.83L/min±5%, can automatically maintain constant flow based on the length resistance of the sampling tube
Display Method800x480mm high-resolution capacitive touch screen display (LCD) Chinese interface, real-time display, previous cycle display, real-time concentration display,
can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling times, status and other parameters, 95% UCL calculation
, can directly display particle concentration(particles/cubic meter), three-level permission management.
Operating Power Maximum Power Consumption:20W (1) Built-in long-life lithium battery: can work continuously for a long time8 hours
External charger can quickly charge, charger power supply is 100V~220V, 50/60Hz
Operating Time8 hours
Counting Mode: Cumulative value, difference value, concentration value
Testing Method: Single, repeat, continuous, calculation, remote
Unit Conversion: Unit can be converted to m³
Particle Size Channels: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0 μm
Sampling Cycle1~9999 (S)Delay Count: 0~9999 (S) Self-cleaning Time: ≤10 (min)
Operating Environment Temperature:0~40℃ (50~104℉)Relative Humidity:20~90%RH, no condensation. Atmospheric Pressure: 86-106KPa
Temperature and Humidity: Optional (1) Temperature: 0~50℃±1℃. (2) Humidity: 0~100%RH±5%
UCL Setting Sampling Points (A): 2~9 points Set Sampling Times per Point: (L) 2~9 times Set Measurement Position: 0-9999
UCL Report Complies with ISO14644-1 2015, GB/T16292, GMP (2010), FS209E
Data StorageCan store 30,000 sets of data (circular buffer) (including particle size, data, environmental data, year, month, day, time, sampling volume,
data port), data is not lost after power failure.
Communication InterfaceAB line, 9600 baud rate, USB interface
Alarm Setting: The instrument has a level alarm function, can alarm for clean room levels 100, 1000, 10000, 300000, 1000000 after exceeding the standard,, simultaneously
Complies with the new version of GMP standards for A/B/C/D four levels of static and dynamic standards measurement, four standard switching measurement
Zero Count Complies with JJF1190-2008 Dust Particle Counter Metrology Calibration Specification and ISO21501-4:2018 less than 1 per five minutes
Overlap Error5%, 3,000,000 particles/cubic foot
Printing Function: Thermal Printer
Weight2kg
Calibration can meet JJF1190-2008 Dust Particle Counter Metrology Calibration Specification, GB/T6167-2007 Dust Particle Counter Performance Test
Method and ISO21501-4: 2018 Cleanroom Light Scattering Dust Particle Counter Calibration Method, traceable to Shanghai Institute of Measurement and Testing Technology
Research Institute can also self-calibrate or use third-party national metrology institutions for calibration, also traceable to the US National Institute of Standards and Technology (NIST).
Our company has passed the national metrology establishment assessment, and the company's standard dust particle counter has also participated in the national dust
particle counter value comparison organized by provincial metrology institutions, and is used as a standard precision particle counter for dust particle counter metrology calibration work by relevant metrology testing institutions.
Standard ConfigurationProfessional isokinetic sampling head, sampling stand with sampling tube, built-in printer, aluminum alloy case, operation manual, factory test report, zero
count filter, charger, sampling stand, printing paper
Dimensions138×180×120 (mm)
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 星源洁净 |
| Spec: | OPC-3016T |
| Stock: | 100 |
| Manufacturer: | |
| Origin: | China / Jiangsu / Suzhoushi |